Fault description

Possible cause

Required measures

Refrigerating performance

Low or no refrigerating perfor-

mance.

Ambient temperature > 25 °C / 77

°F (chap.3.4).

Select cooler place of installation.

Combination of tempera-

ture/humidity values not in the

optimum range (see temperature

humidity diagram, Figure 18).

Select combination of tempera-

ture/humidity values in the opti-

mum range (chap. 12).

Compressor not turned on.

Contact BINDER service.

Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.

Reduce heat load.

Humidity
Humidity fluctuation:

Control accuracy of ± 3 % r.F. is

not reached.

Door gasket defective.

Replace door gasket.

Door opened very frequently.

Open doors less frequently.

Humidity fluctuation, together with

temperature fluctuation >

1 °C with a set-point ca. 3 °C

above ambient temperature.

Place of installation too hot.

Select cooler place of installation

or contact BINDER service.

Low or no dehumidification.

Capillary tube blocked

Contact BINDER service.

Not enough refrigerant.

Icing at the evaporator plates.

Set-point was too long below

ambient temperature.

Defrost the unit (chap. 13).

Condensation at the walls of the

inner chamber.

Combination of tempera-

ture/humidity values not in the

optimum range (see temperature

humidity diagram, Figure 18)

Select combination of tempera-

ture/humidity values in the opti-

mum range (chap. 12).

Set-point was too long below

ambient temperature, icing in the

preheating chamber.

Defrost the unit (chap. 13)

Low humidity and temperature

accuracy

Fan speed has been reduced.

Set fan speed to 100%.

Notification or alarm message

“HUMID SYSTEM” on the control-

ler.

Notification or error of the humidi-

ty system.

See chap. 11.2.

Controller
No unit function

(dark display).

Display mode “Standby” active.

Press any controller key.

Main power switch turned off.

Turn on the main power switch.

No access to menu “User set-

tings”“.

User code incorrect.

Contact BINDER service.

Wrong temperature alarms, dis-

turbance of temperature accuracy Temperature unit changed to °F.

Set temperature unit to °C (chap.

6.4).

Chart recorder function: meas-

ured-value memory cleared; in-

formation lost.

New setting of storage rate.

Change the storage rate ONLY if

the previously registered data are

no longer required (chap. 7).

Controller does not attain set-

points entered in Manual Mode.

Button “EXIT” or “AUTOMATIC”

has been hit: Unit is in Idle Mode.

Change to Manual Mode (chap.

8).

Controller does not attain pro-

gram set-points.

Button “EXIT” or “AUTOMATIC”

has been hit: Unit is in Idle Mode.

Start the program again (chap.

9.9).