Casella CEL CEL-181 User Manual

Page 14

Advertising
background image

Casella USA
17 Old Nashua Road #15
Amherst, NH 03031-2839

9 Calibration Certificate

Instrument Type …………………………………..

Serial No …………………………………………….

Microphone Type …………………………………

Serial No …………………………………………….

Criterion Level (dB) ………………………………

Pa2h (q = 3 vers only) ……………………………..

RMS Time Constant ……………………………..

Amplitude Weighting Function (q) ………………..

Dynamic Range(High) ……………………………

Inst. Overload ……………………………….………

(Low) ……………………………

RMS Overload ……………………………………...


The above noted instrument has been inspected and tested and found to comply with all the
requirements of its relevant specification. The microphone calibration curve for the types CEL-180 is
enclosed with this manual whilst the responses for the other types and details of the other calibration
parameters are retained by the manufacturers under the above-noted serial numbers.

When used with a CEL-182 Dose Meter Calibrator this instrument should be set to a count (on the high
range) of:……………………..


10 ........Modifications and Post Production Developments

Additional Information for US OSHA Regulations

The CEL-181L Personal Noise Dosimeter has the following specification for OSHA classification:

RMS detector characteristic

SLOW

Integration 'q' factor

5 dB (Exchange rate doubling factor)

Measurement range

80-140dB

Criterion level

90 dB

Overload

115 dB RMS continuous

140dB

Instantaneous

Calibration count with CEL-182 9


The relationship between sound level, noise dose count and exposure tine, is as follows for 5 dB
doubling:

L

A

= K + 16.6 log

10

NDC - 16.6 log T


where NDC = % noise dose count and T = exposure time in hours


In the case of dosimeters with criterion levels of 90dB the K = 71.8 and for dosimeters with criterion levels
of 85dB, K = 66.8.
CEL Instruments are manufactured and serviced by:

Casella CEL Inc.

17 Old Nashuia Road #15,

Amherst, NH 03031 USA

PHONE (800) 366-2966

FAX (603) 672-8053

[email protected]


In line with our policy of continual design improvement we reserve the right to amend any of the details in