Measurement ranges:

Range Frequency Display Primary Linearity

Setting Weighting Range Range IEC 651

A LO (low) A

30-100 dB

33-86 dB

A HI (high) A

65-135 dB

68-12 dB

C LO (low) C

30-100 dB

33-86 dB

C HI (high) C

65-135 dB

68-121 dB

Note:

The lower reading is quoted at +10 dB

on noise floor. Upper reading allows

crest factor 5.

Directional response:

Within the limits of IEC 651.

Microphone type:

CEL quarter-inch diameter pre-polarised electret,

typically 10 mV/Pa, permanently attached to the

instrument.

The microphone meets Type 2 requirements for

both free field and random incidence

measurement.

Impedance to be substituted for micro-

phone:

Not possible, microphone permanently attached;

contact CEL for method of measuring electrical

performance.

Calibration reference conditions:

Free field perpendicular incidence,

Ambient temperature: 20

o

C,

Relative humidity: 65%,

Measurement Range: HI (high),

Excitation S

PL

: 114 dB,

Frequency: 1 kHz.

Effect of humidity on accuracy:

Less than

±

0.5 dB over the range 30 to 90% RH

(provided there is no condensation), relative to the

value at 65% RH and 40

o

C.

Storage range: 20 to 95% RH non-condensing.

Effect of temperature on accuracy:

Less than

±

0.5 dB over the range -10 to +40

o

C.

For temperatures between +40 and +50

o

C it is

recommended that the instrument be recalibrated

at the operating temperature with a CEL-282 or

CEL-284/2.

However above 40

o

C, accuracy within the limits

of IEC 651 can be obtained using the following

correction, where t is

o

C:

Corrected reading

= display reading+ 0.05(t-40).

Operating temperature range: -10 to +50

o

C.

Storage range: -20 to + 60

o

C.

Effect of magnetic fields on accuracy:

Negligible at all levels.

Lower than a reading of 30 dB(A) when tested

according to IEC 651 clause 8.4.

Effect of vibration on accuracy:

Less than 3 dB when tested according to IEC 651

clause 8.3.

Effect of operator:

Negligible when instrument held at arm's length

with microphone pointing away from operator.

Frequency weightings:

A- and C-weightings according to ANSI S1.4 Type

2A, and IEC 651 Type 2.

Time weightings:

F (fast):

125 milliseconds,

S (slow):

1 second,

I (impulse): 35 millisecond rise time,

1.5 second decay time.

Digital display:

4 digits,

7 segments,

10 mm high characters,

0.1 dB resolution.

Display updated 2 times per second.

Functions indicated:

S

PL

:

Noise levels dB(A) and dB(C),

M

AX

:

Holds noise readings of F, S, & I, with

no decay,

L

EQ

:

Calculated L

eq

values based on a

sampling rate of 160 samples per

second,

B

ATT

:

Battery voltage,

Low battery level. (Indicated when

battery falls below approximately 3.2

V),

Over load,

Under range.