Manufacturers servicing & warranty arrangements – Casella CEL CEL-269 User Manual

Page 7

Advertising
background image

Over load detection

:

Pre-weighting overload detector, with 2.5 second

dwell time.

Under range detection

:

In S

PL

and M

AX

modes, under range is indicated

when the measured RMS level falls below 65 dB

(HI Ranges) or 30 dB (LO Ranges), there is a 2.5

second dwell time.

In L

EQ

mode, under range is indicated if the

energy in a sample falls below 64.5 dB (HI

ranges) or 29.5 dB (LO ranges).

Otherwise the calculation would include at least

one incorrect sample. This is a permanent

indication that can be removed only by resetting

the instrument.

Start up stabilisation time:

20 seconds.

Output:

Standard 3 pole 3.5 mm coaxial socket with a.c

on pin and ground on sleeve.

AC:

Conditioned signal, 7.25 V RMS for

FSD,

Recommended impedance of load

>39 k

.

Batteries:

4 x IEC AAA Manganese Alkaline type.

Battery life is typically better than 20 hours with

continuous use.

Dimensions:

258 x 70 x 21 mm, (10 x 2.75 x 0.85 inches).

Weight:

200 g, (0.45 lb) excluding batteries.

The Manufacturers reserve the right to change instrument specifications without prior notice.

Manufacturers Servicing & Warranty Arrangements

In order to ensure its rigid conformity with the

requirements of the specification, this instrument

is thoroughly inspected and calibrated prior to

dispatch from the factory. All technical

information for an individual instrument is filed

under the instrument serial number. Therefore,

the serial number should be quoted in any

correspondance concerning the instrument.

The manufacturers undertake to rectify any

defect in the instrument that is directly

attributable to faulty design or assembly, and

which becomes apparent during the warranty

period. In order to take advantage of this

warranty, the instrument must be returned,

carriage paid, to the manufacturer's factory or

accredited agent, where necessary repairs will be

carried out. The warranty period beginning with

the initial date shown on the Certificate of

Compliance is normally 12 months, but with

exceptions on certain specialised components

supplied by other manufacturers which are

warranted for shorter periods.

Some of the specialised components used in this

instrument may be subject to longer guarantees

by their actual manufacturers, and in all such

cases, the benefit of these undertakings will be

passed on to the user. However, CASELLA CEL's

liability is limited to items of their own

manufacture, and they do not accept liability for

any loss resulting from the operation or

interpretation of the results from this equipment.

A comprehensive Instrument Calibration

Maintenance Agreement (ICMA) scheme is

available to extend the initial warranty period of

this instrument. At the end of the first warranty

period, it is recommended that the equipment be

returned to the CASELLA CEL Service and

Re-calibration Department at Bedford, where it

will be inspected and entered into the ICMA

scheme as required. The warranty will then be

extended for the period stated on the individual

schedule. Please contact your local CASELLA

CEL agent for full details of this service.

In addition, CASELLA CEL operates a NAMAS

accredited laboratory No. 0237 which is able to

perform certain acoustic calibrations with

traceability to U.K. National Standards, details on

request.

In the event of a malfunction developing during

the warranty period, the instrument should be

carefully packed and returned either to CASELLA