DIELECTRIC STRENGTH TESTER FOR ISOLATING OILS OT2-60D --------- Page 17
MODEL SPECIFICATIONS AND OPERATING ENVIRONMENT
OT2-60D
Input
230 V ± 10 % , 50/60 Hz, 3 A, single fase.
Output
0-60kVac, 800VA resistive load, between bushings.
Output Termination
Dual Capacitively Graded Bushings.
Oil Test Cells
Available
IEC 156
Oil Test Cells
Under demand
ASTM D1816, ASTM D877
Operating
Temperature
5
°
C a 40
°
C
Distortion
<5%
Meter Accuracy
2% Full Scale.
Kilovoltmeter
3.5 digit LED Escalado 0-60.0 kV AC (RMS)
Dimensions
380mm wide x 360mm deep x 295mm high
Weight
(27kg)
Altitude
100% of rating; Sea-level, up to 1500M. The maximum output
voltage is de-rated 5% above 1500M altitude, 10% above
3600M, and 15% above 4500M.
Storage Temperature
-10
°
C to 60
°
C, (14
°
F to 140
°
F).
Operating
Temperature
5
°
C to 40
°
C (40
°
F to 104
°
F).
Maximum Relative
Humidity
80% up to 31
°
C(88
°
F), decreasing linearly to 50% at 40
°
C(104
°
F)
Installation
Category II
Pollution
Degree 2
Table OT2-60D specifications