CIRCUTOR OT2-60D User Manual

Page 17

Advertising
background image

DIELECTRIC STRENGTH TESTER FOR ISOLATING OILS OT2-60D --------- Page 17

MODEL SPECIFICATIONS AND OPERATING ENVIRONMENT


OT2-60D

Input

230 V ± 10 % , 50/60 Hz, 3 A, single fase.

Output

0-60kVac, 800VA resistive load, between bushings.

Output Termination

Dual Capacitively Graded Bushings.

Oil Test Cells
Available

IEC 156

Oil Test Cells
Under demand

ASTM D1816, ASTM D877

Operating
Temperature

5

°

C a 40

°

C

Distortion

<5%

Meter Accuracy

2% Full Scale.

Kilovoltmeter

3.5 digit LED Escalado 0-60.0 kV AC (RMS)

Dimensions

380mm wide x 360mm deep x 295mm high

Weight

(27kg)

Altitude

100% of rating; Sea-level, up to 1500M. The maximum output
voltage is de-rated 5% above 1500M altitude, 10% above
3600M, and 15% above 4500M.

Storage Temperature

-10

°

C to 60

°

C, (14

°

F to 140

°

F).

Operating
Temperature

5

°

C to 40

°

C (40

°

F to 104

°

F).

Maximum Relative
Humidity

80% up to 31

°

C(88

°

F), decreasing linearly to 50% at 40

°

C(104

°

F)

Installation

Category II

Pollution

Degree 2

Table OT2-60D specifications

Advertising