Beurer TL 80 User Manual

Page 52

Advertising
background image

52

752.773

0113

Irrtum

und

Änderungen

vorbehalten

recommended separation distances between

portable and mobile rf communications equipment and the model Tl80

The models TL80 are intended for use in electromagnetic environment in which radiated RF distur-

bances is controlled. The customer or the user of the model TL80 can help prevent electromagnetic

interference by maintaining a minimum distance between portable and mobile RF communications

equipment (transmitters) and the model TL80 as recommended below, according to the maximum out-

put power of the communications equipment.
rated maximum out-

put power

of transmitter

W

Separation distance according to frequency of transmitter

m

150kHz to 80mHz

d=1.2×P

1/2

80mHz to 800mHz

d=1.2×P

1/2

800mHz to 2,5GHz

d=2.3×P

1/2

0,01

0.12

0.12

0.23

0,1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation

distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-

mitter, where P is the maximum output power rating of the transmitter in watts (W) accordable to the

transmitter manufacturer.

NoTE i At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.

NoTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

Advertising