Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
The FT 65 Multi-function Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user

of the FT 65 Multi-function Thermometer should assure that it is used in such an environment.
IMMuNITY test

IEC 60601 test level

Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floor should be wood, concrete or ceramic tile. If floors

are covered with shythetic material, the relative humidity

should be at least 30%.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV air for input/out-

put lines

Not applicable

Mains power quality should be that of a typical commercial

or hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

Not applicable

Mains power quality should be that of a typical commercial

or hospital environment.

Voltage dips, short interrup-

tions and voltage variations

on power supply input lines

IEC 61000-4-11

<5% U

T

(>95% dip in

U

T

for 0.5 cycle

40% U

T

(60% dip in

U

T

for 5 cycles

70% U

T

(30% dip in

U

T

for 25 cycles

5% U

T

(>95% dip in

U

T

for 5 sec

Not applicable

Mains power quality should be that of a typical commer-

cial or hospital environment. If the use of the FT 65 Multi-

function Thermometer requires continued operation dur-

ing power mains interruptions, it is recommended that the

FT 65 Multi-function Thermometer be powered from an

uninterruptible power supply or a battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels char-

acteristic of a typical location in a typical commercial or

hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.