12. Trouble-shooting

Display

Cause

Solution

The recorded temperature is higher than

1) Ear/Forehead thermometer mode: 109.4 °F (43°C)

2) Object temperature mode: 212 °F (100°C)

Operate the thermometer only between the specified

temperature ranges.

If necessary, clean the sensor tip. In the event of a

repeated error message, contact your retailer or Cus-

tomer Services.

The recorded temperature is lower than

1) Ear/Forehead thermometer mode: 93.2 °F (34°C)

2) Object temperature mode: 32 °F (0°C)
The operating temperature is not in the range from

60.8 to 95 °F (16°C to 35°C).

Operate the thermometer only between the specified

temperature ranges.