Operators Manual

Indicator Does Not Respond When You Step on the Scale

The other load cell is defective.

Check its cable for cuts or flattened

areas or send it back for repair.

Check the defective load cell by

connecting it to the indicator and

verify that it does not weigh.

YES

NO

The other load cell is defective. Check

its cable for cuts or flattened areas and

send it back for repair.

Both load cells are defective. The

indicator is probably OK. Check the

cables for damage and send both

load cells in for repair.

The indicator is defective. Send it in for

repair. Note: Load cells may also be

defective. Check the load cells by

attaching to a good working indicator.

YES

YES

NO

NO

Disconnect one of the load cells and stand over the connected load cell.

Does the indicator display a reading about twice your weight?

Disconnect the first load cell and connect

the other load cell to the same connector

on the indicator. Does the indicator display

a reading about twice your weight?

Perform the paper clip test. Take two paper clips and carefully insert one

into pin 2 and the other in pin 3 of either load cell connector jack on the

indicator. Hold one paper clip in each hand (if the touch it will not damage

the indicator). Does the reading on the indicator jump around?