4 set-up menu – Electro-Chemical Devices (ECD) SMS22 User Manual

Page 21

Advertising
background image

2 S-- 50.00 ppm
Cal 30.0mV/dec

MENU SELECT

▼ (accept Calibration)

Slope has been set at 30.0 mV/decade

2 S-- 50.00 ppm
Cal 30.0 mV/dec

CALIBRATE

▼ (return to Home Screen)

pH Calibration

Screen Displayed

Button Pressed

Ch2 Buffer
Ch2 Setup
Ch2 Status

◄ CALIBRATE ► (enter the calibration menu by pressing
both ◄► simultaneously)

1 pH 7.00 pH
Cal .0 mV

◄ CALIBRATE ►(enter line, zero pt. Cal 1)
Sensor should be in a 7.0 pH solution.

1 pH 7.00 pH
Cal 0.2 mV

MENU SELECT

▼ (accept Calibration when reading is

stable)

1 pH 7.00 pH
Cal 0.2 mV

MENU SELECT

▼ (move to Cal line)

1 pH 7.00 pH
Cal 0.2 mV

MENU SELECT

▼(move to Cal 2, Slope)

Remove and rinse sensor with distilled water

2 pH 10.00 pH
Cal 59.1 mV/pH

◄ CALIBRATE ► (enter Cal Line)
Place Sensor in a pH 10.00 solution

2 pH 10.00 pH
Cal 60.2 mV/pH

Set the pH value using the CALIBRATE

◄ or ► to move

cursor. Use the

▲or▼ to adjust the integer value.

2 pH 10.00 pH
Cal 60.2 mV/pH

MENU SELECT

▼ (accept Calibration)

2 pH 10.00 pH
Cal 60.2 mV/pH

CALIBRATE

▼ (return to Home Screen)

3.3.4 Set-Up Menu
The Setup menu allows the various parameters to be configured for the user’s requirements. Each
Channel has a Set-Up Menu that includes the parameters associated with that specific Channel. The
Graphical Display, 4-20 mA Outputs Timers and Alarm Relays are all configured in the Setup Menu.

Set Up Channel 1 Menu

Screen Displayed

Button Pressed

Ch1 Buffer

Ch1 Setup

Ch1 Status

◄ CALIBRATE ► (enter Set Up Menu)

Plot 4-20 1 Graphical Display

Sample .1 m

MENU SELECT ▼

(the “4-20 1” indicates the displayed value, variable)

Plot 4-20 1

Sample .1 m

MENU SELECT ▼

(the Sample time is 0.1 minutes, variable)

4mA1 .0 ppm OUTPUT SET-UP

20mA1 999.9 ppm

◄ CALIBRATE ► (enter 4mA line)

SMS-22

Page 12