Exchanging electronic insert, Tools, Technical data – Flowserve LRGT16-1 Spector compact Conductivity Transmitter System User Manual

Page 16

Advertising
background image

Inaccurate readings
Fault:

Indicated conductivity higher than measured comparison value.

Remedy:

During commissioning: Reduce temperature coefficient Tk.
During operation: Reduce cell constant.


Fault:

Indicated conductivity lower than measured comparison value.

Remedy:

During commissioning: Increase temperature coefficient Tk.
During operation: Increase cell constant.


Fault:

The measurement result can no longer be adjusted by changing
the cell constant.

Remedy:

Remove equipment and clean measuring surface.

Exchanging electronic insert
1. Undo lid screws and remove both housing lids.

2. Unscrew fixing screws of cover plate and remove the plate.

3. Pull electrode wires from terminal lugs on circuit board. Plug out the lower terminal

strip.

4. Undo fixing screws of electronic insert and remove insert. The electronic insert is

available as spare part (type LRV 1-40).
When ordering please state the version (VS) and material number indicated on
the name plate.

5. Install the new electronic insert in reverse order.

Tools

!

Screwdriver size 1 for cross-recess head screws.

!

Screwdriver size 2.5 for slotted screws, completely insulated to VDE 0680

Technical data

Type code

Type L R G T 1 6 -

1

Conductivity measurement
Closed-loop control
Sensor
Transmitter
Type of measurement (conductimetry = 1)
Pressure rating, PN 40 = 6
Design series
Service pressure
32 bar at 238 °C

16

Advertising