Aligning, removing and mounting the terminal box


Align terminal box
1. Undo cover screws (cross-recess head screws M4)

and remove one terminal cover. The arrow on the
name plate points towards this cover.

2. Loosen fixing nut using 18 (19) mm open-end

spanner and turn terminal box into desired direction
(cable outlet).

3. Re-tighten fixing nut with a torque of 30 Nm.
4. Re-install cover, tighten cover screws, ensuring

correct seat of the cover gasket.

If required you can also remove the terminal box from
the electrode and re-install it afterwards.


Remove terminal box
1. Loosen cover screws (cross-recess head screws M4) and remove one terminal cover.

The arrow on the name plate points towards this cover.

2. Undo fixing screws for cover plate and remove plate.
3. Pull electrode cables from terminal lugs (electronic insert, functional earthing).
4. Loosen fixing nut with 18 (19) mm open-end spanner and remove it.
5. Pull electrode cables through fixing nut.
6. Remove terminal box. Pull all electrode cables through hole for fixing screw.

Install terminal box
1. Pull all electrode cables through hole for fixing screw in terminal box.
2. Place terminal box onto electrode and turn it into the desired direction, taking the cable

outlet into consideration. Ensure correct seat of the gasket between the electrode and
the terminal box.

3. Pull all electrode cables through fixing nut. Screw nut on fixing bolt and tighten with a

torque of 30 Nm.

4. Connect electrode cables according to wiring diagram to terminal lugs (electronic

insert, functional earthing).

5. Screw back cover plate. Install terminal cover and tighten cover screws, ensuring

correct seat of cover gasket.