Tools

Open-end spanner (A. F. 22 mm or A. F. 27 mm)

Torque spanner (A. F. 22 mm or A. F. 27 mm)

Installation

TRG 5-53, TRG 5-54

1. Clean sealing surfaces.

2. Put gasket 13 x 26 onto the sealing surface (choose the material of the gasket

according to the respective application).

3. Screw in temperature sensor and tighten with spanner when cold (torque 50 Nm).

TRG 5-55, TRG 5-56, TRG-5-57, TRG 5-58

1. Remove socket-weld end from temperature sensor.

2. Clean welding surfaces.

3. Arc-weld trap manually (welding process 111 complying to DIN EN 24063) or use

gas-welding process (welding process 3 complying to DIN EN 24063).

4. After welds have cooled down screw in temperature sensor and tighten with

spanner when cold.

Note

To install the temperature sensor in a pipe, weld knee bend to the pipe
such that the measuring tip is immersed in the fluid against the
direction of flow. Fig. 8, Fig. 9

Make sure that the measuring tip of the installed temperature sensor is
always in contact with the fluid. Fig. 3, Fig. 4

The temperature sensors must not be completely insulated. For more
details see examples of installation Fig. 6, Fig. 7