Flowserve Apex 5000 Positioner User Manual

Page 3

Advertising
background image

© 2011, Flowserve Corporation, Provo, UT

B00062a4 (AC-53 MFG) 03/11

Apex 5000 Positioner

Installation, Operation and Maintenance Instructions

CONTROLS

Accord Controls

765 South 100 East

Phone: 801 373 4576

A Unit of Flowserve Corporation

Provo, Utah 84606-6160

Facsimile: 801 489 2591

www.accord-controls.com

Maintenance:

The Apex 5000 positioner is designed for long life and

trouble-free operation. The following steps should be followed

every six months to assure proper operation.

1. Check air supply and associated filtration

equipment. See “Supply Air Requirements.”

2. Make sure arms, bearings, and adjustment screws

move freely. Caution:

Moving parts to check

freedom with supply pressure connected will cause

rotation of cam. Be sure to keep fingers away from

cam pinch points. If parts do not move freely,

disassemble and lubricate with a light, instrument

grade grease (Lubriplate MAG 1 or equivalent).

3. Check for smooth operation of the spool valve. If

it sticks or feels “gummy”, remove it and clean

both spool and valve bore with solvent. Make

sure both parts are clean and dry before

reinstalling. Do NOT apply grease to these parts.

4. Check for air leaks in air supply.

5. Refer to “Troubleshooting” sections if

maintenance does not cure a problem.

Troubleshooting:

Problem

Probable Cause/Solution

Actuator/Valve won’t stroke or

1. Positioner ports are connected to wrong sides of actuator.

goes full stroke with no control:

2. Cam on wrong side (see “Cam Installation”).

3. Stuck arm or spool valve (see “Maintenance”).

4. Input section problem (see “I/P Module Troubleshooting”).

Calibration Shifts:

1. Loose mounting hardware.

2. Loose cam nut.

3. Loose lock screws/nut.

Excessive air consumption (other

1. Spool valve gasket leaking or loose screws.

than normal exhaust of 10-30 scfh):

2. Air leak between positioner/actuator or within actuator.

3. If condition occurs at end of stroke only, calibration is incorrect.

Oscillation or hunting:

1. Air leak between positioner/actuator or within actuator.

2. Oversized spool valve (decrease size).

3. Gain setting too high (see “Gain Adjustment”).

Sluggish or slow response:

1. Undersized spool valve (increase size or add volume boosters).

2. Gain setting too low (see “Gain Adjustment”).

3. Tubing restriction (check supply between positioner/actuator).

4. Tubing too small (increase tubing size).

I/P Module:

General:

All I/P Modules are factory calibrated and should not

require adjustment. However, long-term storage or storage in

temperature extremes may affect calibration. The factory-

installed pressure gauge may be used for calibration, but

accuracy is limited (+/- 3 percent).

Ordering Information:

The Apex 5000 Modular Positioning

System (AMPS) allows easy field installation of four input

options. Specify the following for changeover kits (Kits

include modules, mounting hardware [if necessary],

fasteners, and O-ring seals):

KM50

- Change I/P Module input to 3-15 psi pneumatic input.

KM51

- Change to Model 5100 I/P Module.

KM52

- Change to Model 5200 I/P Module.

KM53

- Change to Model 5300 I/P Module.

KM54

- Change to Model 5400 I/P Module.

KM56

- Change to Model 5600 I/P Module.

Page 3 of 8

Advertising
This manual is related to the following products: