Aspirate and introduce antibody solutions, Incubate – Hoefer PR625 Immunoblot User Manual

Page 6

Advertising
background image

p4

Aspirate and introduce antibody solutions

1

Aspirate excess liquid from the channels through the
numbered holes.
Note: To avoid drying of the membrane, channels
should be loaded within five (5) minutes of aspirating.

2

Pipette antibody solution through the numbered holes.
Press the pipette tip firmly into each hole and inject
the liquid rapidly in a single smooth action until the
channel is filled. Take care not to introduce bubbles.

3

3

Add buffer to all unused channels that cover the
membrane.

number of  

approximate 

model 

channels 

channel volume

4

ImmunoBlot

25

250 µl

ImmunoBlot XL

45

140 µl


Incubate

1

Place the unit on a rocking platform with the channels
aligned in the direction of rocking.
Note: For best results, use a slow rocking speed (5 to
6 tilt cycles per minute). A gyratory shaking platform
is not effective for this purpose.

2

Incubate the unit on the rocking platform for 30 to
60 minutes at room temperature.

Recommendation: Use
a disposable pipette tip
connected to a water
vacuum aspirator for this
purpose.

Important! Take care not
to touch the pipette tip
anywhere on the unit except
in the desired sample hole.

3

See Appendix A, Note 3,
Eliminating bubbles.

4

See Appendix A, Note 4,
Optimal volumes.