Remove and develop the blot, Clean the immunoblot – Hoefer PR625 Immunoblot User Manual

Page 9

Advertising
background image

p7

Recommendation: A 1%
solution of 7X Cleaning
Solution (Flow Laboratories,
50 ml concentrate in 5 liters
of warm water).

Remove and develop the blot

1

Unscrew the ImmunoBlot and separate the two halves.

2

Discard the used sealing pad and wash the membrane
in a tray with 3 to 5 changes of buffer for a total of
10 to 15 minutes.

3

Develop the blot by following instructions from your
Western detection kit (See also Appendix B).

Clean the ImmunoBlot

Clean the unit thoroughly after each use by
performing the following:

1

Rinse the ImmunoBlot unit and washing manifold
under distilled water.

2

Use a non-abrasive, alkaline laboratory cleaning agent
that leaves no residue after rinsing, such as a deter-
gent intended for cleaning tissue culture glassware.
Concentrated cleaning agents should be diluted to
normal strength.

3

Immerse the ImmunoBlot unit and washing manifold
in the cleaning solution and brush gently with a soft
brush taking care not to scratch the interior surfaces.

4

Rinse the ImmunoBlot unit and washing manifold
thoroughly with tap water, followed by distilled water.
To easily rinse the small numbered holes, assemble
the unit and manifold units without a membrane and
sealing pad, and flush with distilled water.

Note: Do not autoclave the
ImmunoBlot unit or washing
manifold.

Do not expose the
ImmunoBlot unit to alcohol or
other organic solvents.