8.2 Technical malfunctions

These malfunctions may occur without an error message being displayed:

Description

Double-Check

Action

Vacuum is not or only slowly released

Is the vacuum released

Wait until the vacuum has been released and

within approx. 30 seconds?

remove object. Switch the furnace off and on again.
If it is still not working, contact your local
Ivoclar Vivadent Service Center.

Microswitch of the furnace head

Is the opening for the operating pin

Remove dirt (foreign objects) from the opening.

opening mechanism does not engage

(item 15 in parts list) obstructed?

The calibration sample assembly

Is the recess obstructed by a foreign object?

Clean the recess with a vacuum cleaner. The

cannot be properly inserted

EP600 should be cold for that purpose.

Loud noises in the press drive

Is the press plunger (56) dirty or damaged?

Clean press plunger or replace it, if necessary

Are the motion rods (81) of the press

Clean motion rods with a dry cloth.

mechanism dirty?

Do not apply grease.

Display incomplete

Run the display test program and contact your
local Ivoclar Vivadent Service Center, if necessary.

Text on the display is difficult to read

Is the contrast set correctly?

Press "Contrast" key (101) until the contrast
is ideal.

Display is not illuminated

Is the fuse (10) for the electronic

Check fuse and replace it, if necessary.

controls OK?

Buzzer does not sound

Is the buzzer switched off (tune 0)?

Select tune (1–9)

Furnace head does not open

Is the fuse (10) OK?

Check fuse and replace it, if necessary.

Was the furnace head opened manually?

Use only the corresponding key to open the
furnace head. Switch furnace off and on again.

Has the vacuum been released?

Is the program still running? Wait until the
program is completed. Switch furnace off and on
again. If it still does not work, contact your local
Ivoclar Vivadent Service Center.

Press plunger fractured in the area

Was the terminal screw of the press

Replace defective press plunger.

of the terminal screw

plunger (57) screwed too tightly?

Press plunger slips out of its holder

Is the press plunger correctly secured?

Tighten the terminal screw of the press
plunger (57).

Press plunger is too long

Was the press plunger of the EP500 with

Use the red press plunger designed for the

a length of 200 mm accidentally used?

EP600 Combi (175 mm).

Vacuum pump does not start working

Is the fuse for the vacuum pump (8)

Check fuse and replace it, if necessary.

defective?

Was the max. power consumption

Use only the vacuum pump recommended

(2.2 A) surpassed?

by Ivoclar Vivadent.

Is the vacuum pump plug correctly

Correctly connect the vacuum pump plug with

inserted?

the furnace base.