6

2.6. INDICATING TRANSMITTERS

An Indicating transmitter in a Divert Control System is identical (both hardware and software) to a standard
transmitter described in the PR-01-S Instruction manual.

A

B

C

D

1

2

3

4

5

6

7

8

9

0

.

-

E N T E R

R E S E T

P R O C E S S T E M P E R A T U R E

22.1 ° C

A

C O N C

T E S T

109.3

In d iv e r t c o n tr o l o p e r a tio n

S ta r t

p r is m

w a s h

R e m o v e d

fr o m d iv

c o n tr o l

C a lib r a te

D is p la y

P R O C E S S IN S T R U M E N T S

K -PT E N T S

4

5

2

3

1

Figure 2.60

The Indicating transmitter.

The display, Figure 2.60, will give the following additional information pertinent to the Divert system:

1. A or B in the upper right corner indicates that the

transmitter is OPERATING in the divert system as re-
fractometer A or B, Figure 2.10. If a or b is displayed
instead, the refractometer is removed from divert con-
trol operation. (If the upper right corner is empty, see
Section 5.2). For A or B the corresponding green
OPERATING light, Figure 2.20, is on, for a or b it is
off.
If you see a ? in this corner, see Section 5.2.

2. A softkey (B) defined by "Remove from divert control"

or alternatively "Restore to divert control". The visible
effects of pressing "Remove" for refractometer A is
that the soft key changes to "Restore", A switches to a,
and the green OPERATING light goes off. Further the
"Remove" soft key disappears from refractometer B,
because B must not be removed if it is the only refrac-
tometer in operation.

Note. Always press "Remove" for a refractometer be-
fore making service or switching the power off.

3. The "Normal operation" message is substituted by

"In divert control operation" or "Removed from di-
vert control". Those messages may change to "Sol-
ids warning" or "Solids alarm" at low solids. An ad-
ditional error message may occur: ** Divert control
fault ** combined with a ? in the upper right corner,
see Section 6.3.

4. The softkey "Display" (D) leads via "System con-

figuration" (C) to "Divert control system" (D) which
gives the following information:

Refractometer difference alarm limit in % solids

Solids warning limit in % solids (default

value 60%)

solids alarm limit in % solids (default value 58%)

Divert rule and SW1/SW2 positions (Section 5.7)

5. The softkey "Calibrate" (C) gives possibilities to set

up for divert control and calibration, see Chapter 5.

INSTRUCTION MANUAL FOR DD-01

DOCUMENT/REVISION: IMD 1/5

Effective: June 20, 2005

Advertising