A-20 API Command Reference

889897

TempScan / MultiScan User’s Manual

*S

- Set Power-Up Configuration

TYPE

System

EXECUTION

Immediate

SYNTAX

*Smode

Specify power-up configuration mode. Valid options for

mode

are:

0

- Power

up under last known configuration (default).

1

- Power up under factory

default configuration.

DESCRIPTION

With the Set Power-Up Configuration (

*S

) command, specifying the mode to be

0

will cause the unit to load its

last known configuration when it is powered on or when the Reset Power-On (

*R

) command is issued. The last

known configuration will be exactly the same as the configuration it had when it was last powered down or a

*R

command was issued. Specifying this mode will insure that the unit will always come up with the configuration
which has most recently been defined. Using this mode is convenient if it is undesirable to reconfigure the unit
every time it is powered up or a

*R

command is issued.

Specifying the mode to be

1

will cause the unit to load the factory default configuration when it is powered on or

when a

*R

command is issued. Using this mode is the same as power-cycling the unit then issuing the Restore

Factory Settings (

*F

) command. For more information on the factory default configuration, see section Power-Up

Configuration in the chapter System Configuration.

Note: Issuing the Set Power-Up Configuration (

*S

) command

*S1

and then power-cycling the unit, or issuing

the Reset Power-On (

*R

) command will cause any previously defined configuration to be irrevocably lost.

However, the calibration factors for the unit and its slaves will remain intact.


EXAMPLE

PRINT#1,"OUTPUT07;Q?T?Y?X"

‘ Retrieve some current configuration settings

PRINT#1,"ENTER07"

‘ Get the settings

INPUT A$

Q1,7,7,7,0 T1,3,0,0 Y100,
1000,0

‘ Screen shows settings

PRINT#1,"OUTPUT07;*S1X"

‘ Configure to power up with factory defaults

PRINT#1, “OUTPUT07;*RX”

‘ Reset the unit

SLEEP 5

‘ Wait for the unit to reset

PRINT#1,"OUTPUT07;Q?T?Y?X"

‘ Retrieve some current configuration settings

PRINT#1,"ENTER07"

‘ Get the settings

INPUT A$

Q1,0,0,0,0 T0,0,0,0 Y0,0,0

‘ Screen shows settings