A-60 API Command Reference

889897

TempScan / MultiScan User’s Manual

X

- Execute

TYPE

System

EXECUTION

Immediate

SYNTAX

X

Execute preceding command string.

DESCRIPTION

Deferred commands are interpreted and processed in the Acquisition Buffer as they are received, but they require
the Execute (X) command to be issued in order to be executed. If multiple system commands are used in the
same string, each use of the command must be followed by the Execute (X) command. Any number of Execute
(X) commands may be inserted into the same command string. However, immediate commands do not require
an Execute (X) command to be processed.

While a command line is being interpreted, the front panel LED indicators will not be updated. These LED
indicators will only be updated when the unit is in a ready state. In order to determine if the unit is in a ready
state, perform a Serial Poll on the Status Byte Register (

STB

) for the Ready Bit (Bit 3).

For more detail on deferred and immediate command types, see section Command Interpretation at the beginning
of this chapter.

EXAMPLE

PRINT#1,"CLEAR07"

‘ Clear the unit

PRINT#1,"OUTPUT07;*C"

‘ Clear the current channel configuration

PRINT#1,"OUTPUT07;C1-16,1 C17,2

C18-32,3"

‘ Send channel configuration

PRINT#1,"OUTPUT07;U4"

‘ Get HLL data for the new channel configuration.

(Note: ERROR LED indicator turns on)

PRINT#1,"OUTPUT07;X"

‘ Instruct the unit to execute the

U4

command. Now the

unit has its new channel configuration

PRINT#1,"OUTPUT07;E?X"

‘ Find out what caused the error

PRINT#1,"ENTER07"

‘ Retrieve error response

INPUT A$

E128

‘ Screen shows

E128

- Command Conflict Error, due to

fact that no channels were configured (

C

is a deferred

command) when the

U4

was executed (

U4

is an

immediate command)

Advertising