Configuration of serial port COM0 and COM1

Baudrate and Framing of the serial ports are configured using the DIL-Switches SW1-1
to SW1-3. COM0 is essential for operation, since the serial time string is read here.

SW1-1

SW1-2

baud rate

off

off

19200

on

off

9600

off

on

4800

on

on

2400

Baudrate of serial ports

SW1-3

framing

off

8N1

on

7E2

Framing of serial ports

Selection of serial timestring

DIP-Switch SW1-4 selects the type of serial time string to be used for synchronization

of the TGV510. The Uni Erlangen/NTP telegram must be selected in case of generating
a IEEE1344 timecode. ( see chapter 'Selection of generated timecode' )

SW1-4

String

off

Standard Meinberg

on

Uni Erlangen/NTP

Selection of serial timestring