Calibrate, Pass/fail – OWON TDS Series User Manual

Page 51

Advertising
background image

5.Advanced User Guidebook

46

●Calibrate

Press the Utility button and select Calibrate in the left menu.
The description of Calibrate Menu is shown as the follows:

Do Self-Calibration

The self-calibration procedure can improve the accuracy of the oscilloscope under
the ambient temperature to the greatest extent. If the change of the ambient
temperature is up to or exceeds 5℃, the self-calibration procedure should be
executed to obtain the highest level of accuracy.

Before performing a self-calibration, disconnect all probes or wires from the input
connector. Press the "Utility" button, select Calibrate in the left menu, select Self
Calibrate
in the bottom menu. A confirmation window will pop up, select OK after
everything is ready.

●Pass/Fail

The Pass/Fail function monitors changes of signals and output pass or fail signals by
comparing the input signal that is within the pre-defined mask.

Press the Utility button and select Pass/Fail in the left menu.

Function Menu

Setting

Description

BackLight

0%~99%

Drag the slider or turn the G knob to adjust
the backlight.

Graticule

Select the grid type

Menu Display

5s~50s, Infinite

Set the disappear time of left and right menu

Gesture

ON, OFF

Enable/disable the touchscreen controls

Function Menu

Description

Self Calibrate

Carry out the self-calibration procedure.

Factory Set

Call out the factory settings.

Touch Screen Calibrate Touchscreen accuracy calibration