OWON TDS Series User Manual

Page 52

Advertising
background image

5.Advanced User Guidebook

47

The description of Pass/Fail Menu is shown as the follows:

Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits of
the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass signal by
built-in and configurable output port. To run the test, read the following steps:

1. Press the Utility button, select Pass/Fail in the left menu.
2. Enable switch on: select Operator in the bottom menu, set Enable as ON in the

right menu.

3. Create rule: select Rule in the bottom menu. Select Source in the right menu;

select the desired source in the left menu. Select Horizontal or Vertical in the
right menu; turn the G knob to set Horizontal or Vertical tolerance. Select Create
in the right menu.

4. Set output type: Select Output in the bottom menu. Choose any one or two of the

options "Pass", "Fail" or "Ring". "Pass" and "Fail" are mutually exclusive
options, which could not be chosen simultaneously.

5. Begin to test: Select Operator in the bottom menu. Select Operator as Start in

the right menu, the test will begin.

6. Save rule: Select Save rule in the bottom menu. Select Save in the right menu to

enter file explorer. The path and file name are shown in the Save To: line in the
bottom. Select the save location, see "To select the path or file in file explorer" on
P43. If you want to customize the file name, select Edit Filename in the right
menu, see "To edit file names" on P43. Select OK Save in the right menu.

7. Recall rule: Select Save rule in the bottom menu. Select Recall in the right menu

to enter file explorer. Select rule file(*.PF), see "To select the path or file in file
explorer
"
on P43. Select the channel that you want to load to in the right menu.

Function Menu

Setting

Description

Operator

Enable

Operator

Control enable switch
Control operate switch

Output

Pass

Fail

Ring

Info

Signal tested corresponds with the rule
Signal tested not correspond with the rule
Ring when it satisfies the rule
Control the display status of info frame on the top left

Rule

Source

Horizontal

Vertical

Create

Select source ①,②,③,④ or Math
Change the Horizontal tolerance value by turning the
G knob
Change the Vertical tolerance value by turning the
G knob
Use the rule set as testing rule

Save rule

Save

Recall

Save the rule to internal/external storage
Recall a rule from internal/external storage as the
testing rule