Troubleshooting – Super Systems SuperOX User Manual

Page 7

Advertising
background image

Super Systems Inc. Page 7 of 14

SuperOX

TM

Operations Manual

Figure 6

Wire designation:

White (Sensor +)

Black (Sensor -)

Green (Thermocouple +)

Red (Thermocouple - )

Troubleshooting

When trouble arises with an oxygen control system, it is important to establish where the

problem is located: the sensor, signal transmission lines, the control instrument, or the

combustion chamber itself. Several simple tests can help to isolate the problem quickly. It is

most important to first understand the nature of the fault. Aside from erratic behavior like

cycling, or failure to stabilize at the setpoint, the most common symptom is non-conformity of

the work pieces to quality assurance specifications.

To evaluate most faults, the recommended tools are:

1. A 3 1/2-digit millivoltmeter with at least 10 M

Ω input impedance and 0 to 1999 mV range;

2. A temperature calibrator; and

3. A simulator to output 0 to 200 millivolts at less than 50 M

Ω output impedance.

Sensor troubleshooting: In order to establish the source of problems in your installation, first

avoid removing the SuperOX

TM

Sensor from the furnace.

All of the following meaningful

questions must be answered while your sensor is at temperature and is exposed to a normal

atmosphere under manual control:

1. Are the connections from the T/C extension wire and sensor cable clean and firmly

attached at the correct sensor and control instrument terminals? Note that the shield

wire in the sensor cable should be connected to ground at the control instrument end

only!

2. Is the sensor impedance less than 50 KΩ at temperatures above 1550°F? Conduct the

test shown in Figure 3 using a shunt resistor of about 100 KΩ. Measure the voltage EC
before shunting, then EM with the shunt in place. Calculate RP. If it exceeds 50 KΩ,

proceed to step 6 below.

3. How quickly does the sensor react to a change in O2 concentration? Read the sensor

millivolts with the controller or the digital meter. Short the sensor for 5 seconds,

remove the short and measure the time required to return to within 1% of the original

reading. If it exceeds 60 seconds, proceed to step 6, below.

Reference

Air

Cooling

Air

Cable

Entry

Advertising