Super Systems SuperOX User Manual

Page 8

Advertising
background image

Super Systems Inc. Page 8 of 14

SuperOX

TM

Operations Manual

4. Is there a leak in the zirconia substrate? To test this property, turn off the reference air

for one minute. Measure the sensor mV as indicated by the controller or a digital

voltmeter. Turn the air back on and measure the mV again. If there is a difference

greater than 5 mV, replace the sensor.

5. Should it be necessary to remove your SuperOX

TM

Sensor from a hot furnace, do so

carefully. UNDER NO CIRCUMSTANCES should it be removed faster than 4" (101.6 mm)

per 5 minutes.

6. If your atmosphere control problem cannot be resolved, our technical support staff is

available Monday through Friday, 7:00 a.m. to 6:00 p.m. Eastern Standard Time. You may

call us at (800) 666-4330.