Simple Dew Digital Dew Point Analyzer Operations Manual

Super Systems Inc.

Page 10 of 18

8.1

Remove the cap of the 11.3% salt solution and install the sensor gland (with the

sensor) into the salt solution. To increase the life of the calibration salts, an effort

should be made to minimize the amount of time that the salt solution is exposed to

the ambient air.

9.0

Allow the sensor to reach equilibrium with the calibration salt.

9.1

Leave the sensor in the calibration salt for a minimum of 24 hours. It is acceptable

to leave the sensor in the salt solution for a longer period of time, even a few days, if

desired.

10.0

Begin the 11.3% (Zero) calibration process

10.1

After leaving the sensor in the salt for at least twenty-four (24) hours, turn the

unit on (if it isn’t on already).

10.2

Simultaneously press the “11.3%” and “Calibration” buttons on the

microprocessor board.

11.0

Verify the 11.3% (Zero) calibration

11.1

Leave the sensor in the 11.3%RH calibration salt

11.2

Record the temperature and the dew point of the sensor. Since there is no

display on the instrument, you will need to measure the voltage from the

microprocessor board and translate that into the appropriate measurement.

11.2.1

To determine the sensor temperature, measure the voltage between pins

5(+) and 9(-) on the microprocessor board. The voltage will be between 0

and 1VDC. Use the chart in Appendix A to determine the temperature in

°F, or use Appendix B to determine the temperature in °C.

11.2.2

Temporarily record the sensor temperature.

11.3

Look up this temperature in Appendix “C” (Determining the Dew Point in °F) or

Appendix “D” (Determining the Dew Point in °C). Appendix C will show the

temperature values in Fahrenheit, and Appendix D will show the temperature values

in Celsius.

11.4

Next to the appropriate temperature, note the number in the corresponding

column titled “11.3%”. This will match the measured dew point, which can be

verified wherever the dew point is displayed.

12.0

After the 11.3% (Zero) calibration has been completed, remove the sensor from the

calibration salt and replace the cap on the salt.