What is dew point, How it works, What is dew point? how it works – Super Systems Simple Dew User Manual

Page 6

Advertising
background image

Simple Dew Digital Dew Point Analyzer Operations Manual

Super Systems Inc.

Page 6 of 18

tip can be cleaned by carefully removing it from the sample block (see Section 2.0 of the Field

Calibration instructions) and rinsing it in isopropyl alcohol. The power should be off while this is

done, and the power should remain off for at least 30 minutes after this procedure to allow all of

the alcohol to completely evaporate.

Water / Moisture Contamination

When a furnace or generator is being started up or cooled down, the resulting gases will

contain unusually high amounts of CO

2

. When these gases cool, moisture will precipitate out

and become condensation inside the sample tubing assembly. Even if the furnace or generator

is operating normally, residual moisture may still be present in the sample tube or plumbing

system. In the same way that the ports are checked for soot (see above) they should be checked

for moisture before attaching the instrument. This is especially important when taking a

sample from a generator, since the sample port is usually preceded by a significant amount of

plumbing. All traces of moisture should be eliminated before attaching the instrument. Failure

to do so will result in erroneous measurements and could result in damage to the analyzer.

The first signs of moisture in the instrument will be visible condensation collecting in the bowl

filter and an unusually high dew point. The upper range of the sensor is +80°F, so if that value

is displayed on the instrument it is probably due to the presence of moisture. If this moisture is

not removed, it will cause the sensor tip to corrode and will eventually require the sensor to be

replaced.

To remove moisture from the instrument, the filter should be removed from the instrument

(since it will probably be wet) and an inert gas such as Nitrogen or Argon should then be flowed

through the instrument for as much time as it takes to dry out. This dry-out time will depend on

the amount of moisture present in the instrument. The condition of the sensor can be

monitored by periodically reading the dew point from the display and watching the value

decrease over time. To test if it is operating properly, verify the ambient dew point against a

web-based weather station that will report the ambient dew point for your area. If the displayed

reading is within three degrees of the reported dew point when the instrument is taken outside,

then all of the moisture has probably been successfully removed. The wet filter and sample

tubing can be re-attached after they have been completely dried out. The filter element will

regain all of its original filtering properties after it has dried out.

To prevent the possibility of moisture damaging the instrument, be sure that the measured dew

point is below ambient levels before it is stored. If necessary, Nitrogen or Argon can be used to

purge the instrument after use.

What Is Dew Point?

Dew point can be defined as the temperature at which the water vapor pressure of the gas

equals the saturated water vapor pressure. In other words, it is the temperature at which

condensation will just begin to occur as the gas is cooled. Dew point and relative humidity are

not the same measurement. Relative humidity is the amount of water vapor in the air compared

to the amount the air could hold if it was totally saturated, and it is expressed as a percentage,

not a temperature. To determine dew point, two main variables are required: Relative Humidity

and Temperature. The Simple Dew measures both variables to compute the dew point.

How It Works

Advertising