Simple Dew Digital Dew Point Analyzer Operations Manual

Super Systems Inc.

Page 4 of 18

Introduction

Thank you for selecting Super Systems Inc. and the Simple Dew as your source for accurate

dew point measurements. The Simple Dew unit is a digital dew point analyzer for standard

range (greater than 0°F or -18°C) measurement. Typical uses of the Simple Dew include

measurement of endothermic atmospheres (with endothermic generators) and

nitrogen/methanol atmospheres, as well as use with plant air systems.

Carefully unpack the Simple Dew - Dew Point Analyzer. If there are any signs of shipping

damage, notify SSi and the shipper immediately.

Specifications

Measurement Range:

0 to +80°F (-18 to +27°C)

Temperature Range:

0 to +120°F (-18 to +49°C)

Power Supply:

115 to 230 VAC (universal input)—for Part No. 13134

24VDC input—for Part No. 13136

Retransmission Output:

0-1 Volt (range is –50 to +80°F)

Size (Closed Case):

15.25” x 9” x 6.25” (approximately)

Weight:

8.2 lbs (3.7 kg)

Warnings

Although it is intended for use in an industrial environment, the Simple Dew is a sensitive piece

of analysis equipment. Care should be taken not to drop the analyzer or to operate it in a

manner inconsistent with its intended use.


Open all sample ports and remove all soot and/or moisture from the lines prior to attaching

the sample tubing for the first time.


The analyzer should be stored at ambient temperature (65-80°F or 18-27°C) for at least two

hours prior to operation.


If the unit is to be returned to SSi for service or any other reason, protect the instrument

with at least four inches of foam or other impact-absorbing material on all sides and place it

in an appropriately sized cardboard box.


This unit is not designed to measure the dew points in corrosive gases, such as Ammonia,

S0

3

, Chlorine, and HCL.


Please read and understand this Product Manual before operating the unit.

Failure to comply with these conditions may cause damage to the unit that will not be covered

under the warranty. Super Systems Inc. is not responsible for damage to this unit caused by

disregard of these warnings, neglect, or misuse.

Startup

The Simple Dew unit was calibrated before it was shipped from Super Systems Inc. You can

begin typical operation as soon as the unit has been allowed to stabilize in a temperature

similar to the temperature in the heat treating department. This stabilization is particularly

important for units that may have been sitting overnight in a delivery vehicle in freezing