Performing a zero calibration, Performing a span calibration, Parts list and internal components – Super Systems Hydrogen Nitrider Analyzer User Manual

Page 7

Advertising
background image

Basic Nitriding Sampling System – Hydrogen Analyzer

Super Systems Inc.

Page 7 of 12

their respective percentages. The more similar to the process gas the calibration gas is, the more

accurate the calibration will be.

Performing a Zero Calibration

On the Sensor Calibration page, select “Zero Hydrogen”. Turn the valve on the side of the enclosure to

the “Calibration Gas” position, and attach the Zero gas to the “Calibration Gas Inlet” port. Begin the flow

of gas at a rate of 1.5 to 2.0 SCFH as measured on the flow meter on the door of the enclosure. The gas

should not be under any pressure other than the amount required to maintain the appropriate flow

amount. The target Value is shown on the screen. For a Zero Calibration, this will be 0.00 (the amount of

Hydrogen in the Zero Gas). The Measured H

2

Value can be seen at the bottom of the screen. When this

value comes to equilibrium, it will not be showing any upward or downward trends, only the slight

oscillation of the readings. This usually occurs in approximately 30 seconds. When the sensor is at

equilibrium, press the green “Calibrate” button to perform the zero calibration. After the Zero

Calibration is complete, turn off the flow of gas and disconnect it from the enclosure.

Performing a Span Calibration

To perform a span calibration, select “Span Hydrogen”, attach the Span Gas to the Calibration Gas Inlet

port, and begin the flow of gas at 1.5 to 2.0 SCFH. The Target Value should be set to the exact amount of

Hydrogen that is in the Span Gas cylinder. Then the same procedure should be followed as the Zero

calibration, with the “Calibrate” button being pressed after the readings reach equilibrium. After the

Span Gas calibration is complete, turn off the flow of gas, disconnect the cylinder from the enclosure, and

restore the valve on the side of the instrument to the “Sample Gas” position. This will re-connect the

sensor to the process gas stream.

Parts List and Internal Components

The following items can be purchased as needed for the Hydrogen Nitrider Analyzer.

Part Number

Description

31604

2 GB SD card

37026

Fan (24VDC, 60mm)

37029

Fan filter assembly

37150

3-way ball valve

37050

High capacity bowl filter

37051

High capacity filter element

34444

Tubing

33097

Circuit breaker

31135

Power supply

A20624

Hydrogen sensor

31296

3.5” touch screen

Full Unit

15357

Hydrogen Nitrider Analyzer

Advertising