14 auto-restart facility – Watson-Marlow 323Dz User Manual

Page 20

Advertising
background image

Watson-Marlow 323Dz User Manual

20

Do not use auto-restart for more than 10 starts per hour.
We recommend remote control where a high number of
starts is required.

14 Auto-restart facility

Auto-restart will re-start the pump after mains power interruptions.

In auto-restart mode, if the pump was previously dispensing a dose, it returns to
the dose start screen and waits for the START key (or remote dose start switch) to
be pressed.

If the pump was previously under manual control, auto-restart returns the pump to
the last manual setting; stopped if the pump was previously stopped, or running if
the pump was previously running.

The default setting is auto-restart off. Without auto-restart the pump displays the
main menu and waits for a control mode to be selected.

To install auto-restart:

Mains power must be available to the pump to engage auto-restart.
Stop the pump. Turn off the mains power switch at the rear of the pump.
Hold down the START key and turn the mains power switch on. The ! symbol
shows on the display.
Start the pump. If the mains supply is interrupted the pump will automatically
restart when the mains power returns.
Auto-restart is retained while the pump is switched off.
To remove auto-restart switch off the mains power at the rear of the pump. Hold
down the STOP key and turn the mains power switch on. The ! symbol will go
out.