iHealth Wireless Blood Pressure Monitor User Manual

Page 10

Advertising
background image

10

The cuff does not
inflate properly or
pressure falls
quickly during test

Review the cuff application
instructions and retest

Irregular heartbeat
(arrhythmia)

It is inappropriate for
people with serious
arrhythmia to use this
monitor.
Check with physician.

The cuff was not
properly applied

Review the cuff application
instructions and retest

Display reads an

abnormal result

The cuff position
was not correct or
it was not properly
tightened.

Review the cuff application
instructions and retest.

Body posture was not
correct during
testing

Review body posture
instructions and retest

Speaking, moving
arm or body, being
angry, excited or
nervous during test

Retest when calm; avoid
speaking or movement during
the test

Bluetooth

connection

unstable

Bluetooth
connection
unsuccessful,
monitor is
abnormal, or strong
electromagnetic
interference is
present

Reset iOS device. Reset
monitor by pressing the
START/STOP button about 10s.
Make sure the monitor and iOS
device are away from other
electrical equipment. Please
see GENERAL SAFETY AND
PRECAUTIONS.

No response

Incorrect operation
or strong
electromagnetic
interference

Press the START/STOP button
about 10 seconds to reset the
device, relaunch app, and
reconnect the iOS device to
the monitor