IEC 60601-1-2:2007 (Medical electrical equipment – Part 1: General
requirements for safety; Collateral Standard-Electromagnetic
compatibility - Requirements and tests);
EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers
- Part 1: General requirements);
EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers
- Part 3: Supplementary requirements for electro-mechanical blood
pressure measuring systems);
ANSI/AAMI SP-10:2002+A1: 2003+A2:2006;
AAMI/ANSI 80601-2-30: 2009/IEC 80601-2-30: 2009+Cor.2010/EN
80601-2-30:2010(Medical electrical equipment –Part 2-30: Particular
requirements for the basic safety and essential performance of
automated non-invasive sphygmomanometers).

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration - electromagnetic emissions

The Wireless Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below.
The customer or the user of the MONITOR should assure that it is used in such an
environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions
CISPR 11

Group 1

The Wireless Blood Pressure Monitor uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.

RF emissions
CISPR 11

Class B

The MONITOR is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.

Harmonic
emissions
IEC 61000-3-2

Class A

Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3

Complies

Table 2

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration - electromagnetic immunity

The Wireless Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Wireless Blood Pressure
Monitor should assure that it is used in such an environment.

IMMUNITY

test

IEC 60601test

level

Compliance level

Electromagnetic environment -

guidance

Electrostat

± 6 kV contact

± 6 kV contact

Floors should be wood,