Operation:

– Change to the second operating level and select the

function group for the Data Logger.

– Set the required number of FILES in FILE NUMBER:

SET #01 = 1 FILE, SET #02 = 2 FILES, etc.

– Simultaneously press M when the required number

is displayed. The memory is correspondingly set up.

You can store the measurement results. The Data Log-
ger has the following presettings:

FILE MODE:

THICKNESS

MEASURE:

0 TO 1st

GATE LOGIC:

POSITIVE

COPY MODE:

DATA LOGGER

Selecting a FILE

If the instrument was switched off or if the Data Logger
was selected, then the FILE number last selected is
automatically active. In order to select another number:

Operation:

– Change to the second operating level and select the

function group of the Data Logger.

– Set the required number in FILE NUMBER.

The active status of each FILE number is displayed:

EMPTY:
File is empty.

ACTIVE:
File contains at least one reading.

FULL:
All measurement locations in this FILE are allocated.

ALL CLEAR:
You must first enter the number of FILES.

Operation

Storing measured values - Data Logger

5-36

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Operation:

– Change to the second operating level and select the

function group for the Data Logger.

– Set the required number of FILES in FILE NUMBER:

SET #01 = 1 FILE, SET #02 = 2 FILES, etc.

– Simultaneously press M when the required number

is displayed. The memory is correspondingly set up.

You can store the measurement results. The Data Log-
ger has the following presettings:

FILE MODE:

THICKNESS

MEASURE:

0 TO 1st

GATE LOGIC:

POSITIVE

COPY MODE:

DATA LOGGER

Selecting a FILE

If the instrument was switched off or if the Data Logger
was selected, then the FILE number last selected is
automatically active. In order to select another number:

Operation:

– Change to the second operating level and select the

function group of the Data Logger.

– Set the required number in FILE NUMBER.

The active status of each FILE number is displayed:

EMPTY:
File is empty.

ACTIVE:
File contains at least one reading.

FULL:
All measurement locations in this FILE are allocated.

ALL CLEAR:
You must first enter the number of FILES.

Operation

Storing measured values - Data Logger