5-46

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

GATE LOGIC Setting the gate logic

In this function you can activate the gate and select
the criteria for triggering the gate alarm. The following
settings are available:

• OFF: The gate is disabled. The gate bar is not

displayed.

• POSITIVE: The gate is enabled. When the signal

exceeds the threshold that was preset in a-THRESH,
the two LEDs will flash alternately.

• NEGATIVE: The gate is enabled. When the signal

falls below the threshold that was preset in a-
THRESH
, the two LEDs will flash alternately.

• MEASURE: The gate is enabled. However, there is

no alarm with threshold violation.

H Note:

You can select either POSITIVE, NEGATIVE or MEA-
SURE for digital thickness measurements.

A

Attention:

Both gates only evaluate within the visible part of the
display range. If there are gates or parts of gates out-
side the visible part of the display range, then a gate
evaluation will not be made there.

Operation:

– Highlight the center function group in the left function

table with I.

– With M select the required gate setting.

Operation

Configuring the USN 52R/USN 52L

5-46

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

GATE LOGIC Setting the gate logic

In this function you can activate the gate and select
the criteria for triggering the gate alarm. The following
settings are available:

• OFF: The gate is disabled. The gate bar is not

displayed.

• POSITIVE: The gate is enabled. When the signal

exceeds the threshold that was preset in a-THRESH,
the two LEDs will flash alternately.

• NEGATIVE: The gate is enabled. When the signal

falls below the threshold that was preset in a-
THRESH
, the two LEDs will flash alternately.

• MEASURE: The gate is enabled. However, there is

no alarm with threshold violation.

H Note:

You can select either POSITIVE, NEGATIVE or MEA-
SURE for digital thickness measurements.

A

Attention:

Both gates only evaluate within the visible part of the
display range. If there are gates or parts of gates out-
side the visible part of the display range, then a gate
evaluation will not be made there.

Operation:

– Highlight the center function group in the left function

table with I.

– With M select the required gate setting.