Instruction Manual

748374-F
March 2003

1-4 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 TO2

1-5 SPECIFICATIONS

a. General

Measurement Species................... Trace Oxygen

Ranges .......................................... 0 to 100 ppm (output scalable down to 0-2 ppm fullscale)

Accuracy........................................

±3% of reading or ±0.02% of range (except for ranges ≤ 100 ppm:
±3% of reading or ±0.05% of range)

Sensitivity ...................................... <10 ppb Oxygen

Noise ............................................. 1% of fullscale, peak to peak

Linearity .........................................

±1% of fullscale

Response Time ............................. Typically 90% in less than 20 seconds

Zero Drift........................................

≤±1% of fullscale/24 hours at constant temperature

Span Drift.......................................

≤±1% of fullscale/24 hours at constant temperature

Effect of Temperature.................... 0.32% of reading per

°F from 70°F

(0.58% of reading per

°C from 21°C)

Effect of Flow.................................

≤2% of reading for a flow change of ±250 cc/min (0.5 SCFH)

Operating Temperature ................. 32

°F to 113°F (0°C to 45°C)

Power Requirements..................... +24 VDC

±5%, 10 W max.

Ripple and Noise: <100 mV peak to peak
Line and Load Regulations: <

±1%

b. Sample

Sample .......................................... Non-flammable (below 100% of the LEL)

Flow Rate ...................................... 0.5 to 1.5 L/min.

Supply Pressure ............................ 1027 to 1082 hPa - absolute (0.2 to 1.0 psig)

Temperature .................................. 32

°F to 113°F (0°C to 45°C)

Particulates.................................... filtered to <0.1 mg/L; non-condensing at ambient temperature

Sample Humidity ........................... non-condensing at ambient temperatures

c. Physical

Materials in contact with sample ... Stainless steel, Teflon, Delrin, neoprene

Dimensions.................................... See Figure 2-2. Trace Oxygen Analyzer Outline and Mounting

Dimensions on page 2-3

Weight ........................................... 6.8 kg (15 lbs.)

Mounting........................................ Horizontal, external to Platform or custom installed in a panel

Case Classification ........................ General Purpose for installation in weather protected area

Max. Separation from Platform...... 1600 m (1 mile)

d. Gas

Connections

Sample In ...................................... 1/4 inch O.D. tube fitting

Sample Out.................................... 1/4 inch O.D. tube fitting

See the Preface Section of the Platform manual for specifications regarding Platform related components.