0 startup and operation, 1 overview, 2 displays – Emerson ROSEMOUNT NGA2000 TO2 User Manual

Page 23: A. run mode display, B. menu displays, 1 run mode display, 2 main menu, Ppm o, Instruction manual

Advertising
background image

Instruction Manual

748374-F

March 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-1

Model NGA2000 TO2

SECTION 3

STARTUP AND OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should perform
the leak test procedure outlined in Section 2.

For the remainder of this section, Analyzer
Module interconnection with a Platform or some
interfacing component will be assumed. Display
and Keypad information refers to that which the
user can expect to see and do with regard to the
Front Panel of the Platform.

(For a complete description of Platform Front
Panel controls and indicators, see the Platform
instruction manual.

3-2 DISPLAYS

Three kinds of Display screens are available to
the user:

• Run

Mode

• Menu

• Help

a.

Run Mode Display

The Run Mode is the normal mode of op-
eration. In this mode, the Display will show
the current gas measurement, the compo-
nent of interest, the current operations of
the softkeys, and several graphics: a bar
representing the displayed concentration as
a percent of fullscale and up to four lines
showing user selectable secondary parame-
ters from either the Analyzer Module or any
IO module bound to it. See the Platform
manual for information as to how to select
these.

If more than one Analyzer Module is con-
nected to the system, an additional Run
Mode display will show as many as four
(five for version 2.3 and later) gas meas-
urements on screen.

Figure 3-1. Run Mode Display

b. Menu

Displays

The Menu structure enables the user to ac-
cess data and functions, and put informa-
tion onto the network. From the Run Mode
display, press the MENUS softkey to gain
access to the Main Menu.












Figure 3-2. Main Menu

The Main Menu is subdivided into three lev-
els of control based generally on which per-
sonnel is likely to use it: Basic Controls -
Operators, Expert Controls and set up -
System Engineers, and Technical level con-
figuration
- Analyzer technicians. Many lay-
ers of the menu structure are described at
appropriate places throughout this manual.

TO2 1

0 Range 3

50

Sensor current: 1.4403 mA

19.4

ppm O

2

Sensor temperature: 26.3 C

Temperature current: 2.01 mA

Sample Flow: 511 ml/min

DISPLAY

PARMS.

MENU

NEXT

INFO

Main Menu

Basic controls . . .

Expert controls and set up . . .
(Operational configuration)

Technical level configuration . . .
(Diagnostic and manufacturing/service)

DISPLAY

PARMS.

NEXT

EXT LOCK

INFO