ALARM / INDICATORS / TROUBLESHOOTING

Green

Ready for use

Amber Caution - Prompt response by the OPERATOR is required

Red

Warning - Immediate response by the OPERATOR is required

The Audible Alarm and LED Indicators are listed in the chart.

Visual Indicator/

Problem

Audible

Alarm

Probable Cause

Remedy

The last LED on

the internal Battery

Level Indicator turns

red

Repetitive beep Internal battery has

dropped below 15%

charge remaining

Connect the POC to the AC Power Adapter,

DC Power Adapter, or external battery pack

The last LED on

the internal Battery

Level Indicator

flashes red

Repetitive beep

Internal battery has

dropped below 5%

charge remaining

1. Connect the POC to the AC Power

Adapter, DC Power Adapter, or external

battery pack

2. Change to alternate oxygen source if

power source is not available

Amber alarm

indicator is

illuminated and

setting number is

not flashing during

inhalation to signal a

detected breath

Repetitive beep No breath detected for

1 minute

1. Verify the cannula is not kinked

2. Verify the cannula is properly connected

to oxygen outlet

3. Verify the cannula is properly positioned

in the nose

Red alarm indicator

is illuminated and

setting number is

not flashing during

inhalation to signal a

detected breath

Repetitive beep

No breath detected for

2 minutes

1. Verify the cannula is not kinked

2. Verify the cannula is properly connected

to oxygen outlet

3. Verify the cannula is properly positioned

in the nose

4. If alarm persists change to alternate

oxygen source and contact your

Equipment Provider

Amber alarm

indicator is

illuminated

Repetitive beep Breath rate exceeding

35 breaths / minute

Reduce level of activity to allow breath rate

to slow

Red alarm indicator

is illuminated

Repetitive beep Breath rate exceeding

35 breaths / minute

for longer than 1 minute

1. Reduce level of activity to allow breath

rate to slow

2. If alarm persists, change to alternate

oxygen source and contact your

Equipment Provider

Red alarm indicator

is illuminated, unit

is not running and

setting numbers 1 &

2 are flashing

Repetitive beep

Abnormal compressor

operation detected

1. Turn the POC to the “OFF” position

2. Turn the POC back “ON” to prescribed

setting

3. If alarm persists, change to alternate

oxygen source and contact your

Equipment Provider

Red Alarm and 2 &

3 Setting Numbers

Illuminated and

device Shuts Down

Single Beep

Cannula Occluded/

Kinked

1. Ensure the Cannula and/or Outlet are

Not Blocked

2. Turn the POC to the “OFF” position for at

least 15 seconds.

3. Turn the POC back “ON” to prescribed

Setting

4. If Alarm persists, change to alternate

Oxygen source and Contact your

Equipment Provider.

All other Alarm

Conditions

Unknown

1. Switch to

another source

of Oxygen.

Unknown

1. Switch to another

source of Oxygen.

1. Switch to another source of Oxygen.

2. Contact Equipment Provider.