Specifications (continued) – Precision Medical PM4130 EasyPulse POC3 User Manual

Page 8

Advertising
background image

7

Portable Oxygen Concentrator

SPECIFICATIONS (continued)

Oxygen Minute Volume*: Setting 1:

240 +/-15% mL/min

Setting 2:

380 +/-15% mL/min

Setting 3:

520 +/-15% mL/min

Oxygen Concentration: 87% to 95% on all settings
Mode of Delivery:

Pulsed Dose only

Trigger Sensitivity:

0.50 cmH

2

O maximum

Trigger Delay:

0.11 seconds typical

Power:

AC Power Adapter:

100-240 VAC (1.0 A @ 120 VAC)

DC Power Adapter:

11-16 VDC (5.0 A @ 12 VDC)

Battery Duration:

Setting 1:

5.5 hrs

(Approximate)

Setting 2:

4.0 hrs

Setting 3:

3.0 hrs

Times may vary based on atmospheric conditions and battery age.
Sound (@ Setting 2):

39 dBA

Cannula: Maximum 7 ft (2.1 m) long standard adult single lumen nasal cannula.
Equipment Classification:
Class II

- with respect to protection from electrical shock

Type BF

- degree of protection against electrical shock

IP21

- Protected from touch by fingers and objects greater

than 12 millimeters. \ Protected from condensation.

Volatile Organic Compound (VOC) and Particulate Requirements:
The oxygen delivered from the EasyPulse POC meets the following requirements for

particulate levels, VOC levels, carbon monoxide levels, carbon dioxide levels and

ozone levels.

EPA PM 2.5

Particulate Matter

ASTM D5466

Ozone Levels

21 CFR 801.415

Carbon Monoxide Levels

EPA NAAQS - Carbon Monoxide

Carbon Dioxide Levels

OSHA Permissible Exposure Limits

Standard Test Method for Determination of Volatile Organic Chemicals in

Atmospheres (Canister Sampling Methodology)

* At Sea level

Specifications are subject to change without prior notice.