Duration trigger (only for ds1000d series) – RIGOL DS1000D User Manual

Page 77

Advertising
background image

RIGOL

© 2008 RIGOL Technologies, Inc.

User’s Guide for DS1000E, DS1000D Series

2-47

Duration Trigger (Only for DS1000D Series)

Duration Trigger Mode: Trigger in appointed time when code terms are satisfied.
Press Trigger MENU→Mode→Duration to display the duration trigger menu as
follows.


Figure 2- 61 Table 2- 37 The Duration Trigger Menu (Page 1/2)

Menu

Settings

Comments

Select

D15-D0

Choose digital channel for Duration Trigger

Code

H
L
X

High
Low
Ignore

Qualifier

<
>
=

Set time limit terms


Figure 2- 62 Table 2- 38 The Duration Trigger Menu (Page 2/2)

Menu

Settings

Comments

Time

<Time
Setting>

Set duration and limit symbol time

Sweep

Auto
Normal
Single

Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop

Set Up

To go to set up menu, see table 2-40


Key Points:
1. H (High):
Logic high: voltage is higher than threshold setting.
2. L (Low): Logic low: voltage is lower than threshold setting.
3. X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t

be triggered.

4. Qualifier: A timer begins when code terms are satisfied. Duration trigger

occurred in the time set by the qualifier.

Advertising
This manual is related to the following products: