Changing humidifier lamps, Figure 26 infrared humidifier lamps, Autoflush infrared humidifier cleaning system – Emerson Liebert Challenger With Liebert iCOM Control 3000/ITR User Manual

Page 57: Autoflush operation, Notice

Advertising
background image

Component Operation and Maintenance

51

Changing Humidifier Lamps

1. Open disconnect switch.
2. Open front panel.
3. Remove screws securing line voltage compartment cover, then remove the cover.
4. In line voltage compartment, disconnect one end of the purple jumpers, then locate the burned out

bulb with a continuity meter.

5. Remove humidifier pan. Refer to Removing the Pan on page 50.
6. Remove lamp brackets (2) under lamps.

Figure 26 Infrared humidifier lamps

7. Loosen two screws securing bulb wires to junction block.
8. Pull bulb straight down.
9. Replace bulb. Wrap wires once loosely around bulb. This will support the bulb and also allow for

thermal expansion. Make sure lamp wires are secure in the junction block.

NOTICE

Risk of oily deposits. Can shorten component life.
Do not touch the quartz lamps with your bare hands. Oily deposits such as fingerprints will

severely shorten bulb life. Use clean cotton gloves at all times.

10. Reverse Steps 1 through 6 to reassemble.

Autoflush Infrared Humidifier Cleaning System

NOTICE

Risk of improper water pressure. Can cause improper component operation.
To operate properly, the Autoflush Humidifier requires a water source that can deliver at

least 1 gpm (0.063 l/s) with a minimum pressure of 20 psig (138 kPa).

The autoflush system will periodically flush the humidifier pan with water to prevent the buildup of

water minerals due to saturation. Because water conditions vary, the amount of water flushing

through the system may be programmed to match local needs.
Water amounts between 110% and 500% of the amount needed for humidification may be selected.

Operation of the flushing system is then automatic and no further adjustments need to be made.

Autoflush Operation

The operation of the autoflush is divided into four steps, beginning with a call for humidification.
1. If the humidifier has not been activated for over 30 hours, the autoflush will flow water into the

pan for about 30 seconds. This will provide a minimum amount of water in the pan and prevent

heat damage to the humidifier pan. Humidifier lamps are Off.

2. If the humidifier has been activated within the last 30 hours, Step 1 is bypassed. The autoflush

will flow water into the pan for about 4 minutes. The humidifier lamps are On and the humidifier

is operational during this period. When the pan is filled (the fill cycle has timed out), the water

make-up valve is closed.

3. The water make-up valve remains Off and the humidifier lamps are On for a maximum of

9-1/2 minutes.

4. After the 9-1/2 minute delay, the autoflush adds water to the pan to replenish the water used in

humidification and flush the pan of mineral solids. This amount of water is adjustable from 110%

to 500% in increments of 10%. At the end of this cycle, the make-up valve is closed. Steps 3 and 4

repeat as long as humidification is required.

Humidifier Lamps

Advertising