Using the t emporal scanner on an infant, Temping baby in bassinette, open crib, or with mom – Exergen TEMPORALSCANNER TAT-5000 User Manual

Page 11

Advertising
background image

Questions? Please call us at 800-422-3006

Using the T

emporal

Scanner

on an Infant

Temping Baby in Bassinette, Open Crib, or with Mom

Instrument should be in same temperature environment as the
baby for approximately 20 minutes.

Measurement site must be exposed.

One measurement, preferably at the TA, is all that is required.

Preferred site is the temporal

artery area. In this case,

behind the ear could be

alternate site, as both are

exposed.

Temporal artery area is

the only option in this

case, as the neck area is

not exposed.

Temperature at the Temporal Artery Area

1)

Gently touch probe to center of forehead.

Depress red button and keep depressed.

2)

Slide probe over the TA area into hairline.

If more convenient, slide from hairline towards center

of forehead.

3)

Release button, remove from head, and record.

Temperature Behind the Ear

1)

Gently nestle probe on neck behind the ear.

Depress red button and keep depressed.

2)

Maintain skin contact until numbers stop.

3)

Release button, remove from head, and record.

9

818528r5:818528r5.qxd 4/24/2008 11:04 AM Page 11