Determining a Fever Threshold

Determining a Fever Threshold for Temporal Artery Thermometry

Threshold Defining Fever

• A threshold for defining fever is the temperature level above which false

positives due to normal variations in temperature, including range of normal
mean + circadian effects + other effects (metabolic, ovulation, etc.) are unlikely.

Threshold for Fever Workup

• Not all fevers require a fever workup. A fever workup is an early management

tool in assessment of the likelihood of septicemia or bacteremia, and initiated
whenever an infectious source is suspected. The level of temperature triggering
such that investigatory workup is sufficiently high to avoid false positives
resulting in unnecessary discomfort and expense for the patient, but low
enough for early identification and intervention.

Primary Points

• Temperatures measured with temporal artery thermometry may be higher

than normally seen with other clinical methods, and therefore require an
adjustment in both protocol and perception.

• No one value can apply to every temperature measurement site. Note old

rule of thumb: Rectal temperature is ~1°F higher than oral temperature and
~2°F higher than axillary temperature.

• Recommended threshold for fever workup using arterial temperature assessment

is a single temperature >101.8°F, or a temperature >101.2°F sustained for
more than 1 hour.

• Adjustment of ~1°F is necessary to raise the temperature level normally

mandated for fever workups to prevent false positives, unnecessary cultures and
blood tests, etc.

Physician Recommended Guidelines for Fever Workup

1

Temperature Site

Fever Workup

Recommendation

Core &

Temporal Artery

Single value

>101.8

Sustained values

(>1h) >101.2

Oral & Temporal

Artery in Oral

Calibration

Single value >101

Sustained values

(>1h) >100.4

Axillary

Single value >99

1

Source on file at Exergen Corporation