Running self test, Run all tests, Running self test 163 – HP OMNIBER 718 HP 37718A User Manual

Page 173

Advertising
background image

163

Selecting and Using "Other" Features

Running Self Test

Running Self Test

Description

Before using the HP 37718A to make measurements, you can run Self
Test ALL TESTS to ascertain the integrity of the HP 37718A. These
tests take between at least 1 hour to complete depending on the options
fitted.
Alternatively you can run Confidence Tests which only takes 2 to 3
minutes to complete. This is not a full verification but performs BER
measurements with internal and external loopbacks fitted.

HOW TO:

Run ALL TESTS

1

Choose TEST TYPE

on the

display.

2

Insert a formatted disk into the instrument disk drive.

3

Make the loopback connections listed below:
Connect Transmit module 75

OUT to Receive module 75

IN

Connect Transmit module 100/120

OUT to Receive module 100/

120

IN

Connect Transmit module 75

MUX to Receive module 75

DEMUX

Connect Transmit module 100/120

MUX to Receive module 100/

120

DEMUX

Connect Multirate Analyser IN to OUT
Connect Optical OUT to Optical IN via a 15 dB attenuator.

ALL TESTS

OTHER

SELF TEST

Advertising