Setting jitter receive interface, Setting jitter receive interface 18 – HP OMNIBER 718 HP 37718A User Manual

Page 28

Advertising
background image

18

Setting the Interfaces

Setting Jitter Receive Interface

Setting Jitter Receive Interface

Description:

Jitter and error measurements are made simultaneously when a jitter
option is fitted. The measurements are made on the normal input to the
DSn or SONET receiver and the interface selections are the normal
Receiver selections. The jitter receive interface is selected with

or

MEASUREMENT TYPE

.

The choices made on the jitter receive interface determine the jitter
measurement range, the threshold level for determining a jitter hit and
which filters are used in the jitter measurement.

HOW TO:

1

Choose MEASUREMENT TYPE

.

2

Choose the RECEIVER RANGE - the jitter measurement range.

3

Choose the HIT THRESHOLD level - if the received jitter exceeds the
value chosen a jitter hit is recorded.

4

Choose the FILTER you wish to include in the peak to peak and RMS
jitter measurement.

RECEIVE

PDH/DSn JITTER

RECEIVE

SONET JITTER

JITTER

JITTER

Advertising