White Calibration for Reflectives

The Flextight scanner uses a white reference to make sure that all

elements in the scanner's CCD react consistently to light brightnesses.

The scanner has been calibrated from the factory, and the calibration

data is stored in the scanner’s flash PROM.

As the scanner ages, the properties of the light tubes may change

slightly. Therefore, you must redo the white calibration every six months

or so and immediately after changing the light tubes.

Use procedure below to make White Calibration:

1. Find the white calibration sheet that came with your scanner. It is a

square white piece of paper (219 x 219 mm (8.6" x 8.6")).

If you have lost the white calibration sheet, then you can use a flat

(non-textured), perfectly white piece of paper of the same

dimensions.

2. If you have used the reflective holder before, make sure that there

are no finger prints or other marks on the plastic layer. Replace the

layer with a new one if necessary.

(Please note that the A4/letter-size reflective holder in not included

in the standard delivery).

3. Align the top edge of the sheet with the dotted line near the top of

the original holder (see Figure 8 opposite). Align each side of the

target with the edges of the original holder.

4. Select White Calibration from the Maintenance window. The

External White Calibration

window appears.

User's Guide - Flextight 646

Hardware Reference - Flextight 646

22

Calibrating the Scanner