Para 4 – HP 3325A User Manual

Page 4

Advertising
background image

TB 9-6625-2049-35

4

Table 1. Calibration Description - Continued

Test instrument parameters

Performance specifications

Amplitude accuracy with dc

offset and no attenuation
(range 1) into 50

load

Tolerance relative to

programmed amplitude

Sine wave: .001 Hz to 100 kHz

±

0.3 dB

Square: .001 Hz to 100 kHz

±

3%

Triangle: .001 Hz to 2 kHz

2 kHz to 10 kHz

±

4%

±

6%

Ramps: .001 Hz to 500 Hz

500 Hz to 10 kHz

±

4%

±

11%

Function and frequency range

Tolerance relative to

programmed amplitude

Attenuator accuracy (these

errors are additive with the

amplitude accuracy errors)

.001 Hz to 100 kHz

Attenuator ranges 2 through 8

±

0.1 dB

100 kHz to 10 MHz

Attenuator ranges 2 through 8

±

0.2 dB

10 MHz to 20 MHz

Attenuator ranges 2 through 4

Attenuator ranges 5 through 8

0.2 dB

±

0.5 dB

Amplitude output

Range: 1.000 mV to 10.00 V p-p

Amplitude (option 002)

(high voltage output)

Range: 4 mV to 40 V p-p (>500

)

Accuracy: +2% of full output for each range at 2 kHz

Flatness: +10% of programmed amplitude

Dc offset

Range: +5 V dc

Accuracy: +0.4% of full peak output for each attenuator range

2

Dc offset (option 002)

Range: +20 V dc

Accuracy: + (1% +25 mV) of full output for each attenuator range

Dc plus ac

Range: <1 MHz

Accuracy: +1.2%

Ramps: +2.4%

Range: >1 MHz

Accuracy: +3%

1

Not verified below 50 Hz.

2

Except lowest attenuator range where accuracy is

±

20

µ

V.

SECTION II

EQUIPMENT REQUIREMENTS

4. Equipment Required. Table 2. identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The
item selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.