Users manual – Fluke 561s User Manual

Page 16

Advertising
background image

561

Users Manual

12

Wrap Thermocouple Probe (model-specific)

Type..............................................................................Type K with miniconnector and

Velcro strap, ASTM E230-03

Standard Tolerance

Measurement Range ....................................................0 °C to 100 °C (32 °F to 212 °F)
Accuracy .......................................................................

±

2.2 °C (4.0 °F)

Total Length..................................................................505 mm (20 in) cable terminated with

a Type K thermocouple inside a
495 mm (19.5 in) nylon Velcro cuff

Bead Thermocouple Probe (model-specific)

Type..............................................................................Type K with miniconnector
Measurement Range .................................................... -40 °C to 260 °C (-40 °F to 500 °F)
Accuracy .......................................................................

±

1.1 °C (2.0 °F) from 0 °C to 260 °C

(32 °F to 500 °F). Typically within
1.1 °C
(2.0 °F) from -40 °C to 0 °C

(-40 °F to 32 °F)

Cable Length ................................................................1 m (40 in) terminated with Type K

thermocouple beads

Electrical

Power Supply................................................................2 AA Batteries (alkaline or NiCD)
Power Consumption......................................................At least 12 hours battery life

Physical

Weight...........................................................................0.322 kg (0.7099 lb)
Size...............................................................................17.69 cm (6.965 in) x 16.36 cm

(6.441 in) x 5.18 cm (2.039 in)

Environmental

Operating Temperature Range .....................................0

°

C to 50

°

C (32

°

F to 120

°

F)

Relative Humidity ..........................................................0 to 90 %, noncondensing up to

30

°

C (86

°

F)

Storage Temperature....................................................-20

°

C to 65

°

C (-4

°

F to 150

°

F)

Optional Accessories ....................................................Soft Case

Advertising