Fluke 434 User Manual

Page 112

Advertising
background image

Fluke 433/434

Users Manual

21-10

PF, DPF, COS

Φ:

Measurement Measurement

Range Resolution Accuracy

(excl. Clamp error)

Power Factor

0.00 ... 1.00

0.01

± 0.03

Displacement Power
Factor

0.00 ... 1.00

0.01 ±

0.03

COS

Φ

0.00 ... 1.00

0.01

± 0.03


Harmonic Power (Watts only)

Settings Range

Additional

Information

Harmonic selection (n):

DC, 1 ... 50

Grouping: Harmonic Groups

Amplitude Reference

total Power / fund. Power Used for Relative Amplitude

THD

% total / % fundamental

based on H1 ... H40

Measurement: Measurement

Range Resolution

Accuracy

(excl.

Clamp error)

Relative Amplitude

0.0 ... 100.0 %

0.1 %

± n x 2 %

Absolute Amplitude
0.1 mV/A ... 1 V/A
V

∗1 ... V∗1000

as indicated under W,
VA, VAR ranges

as indicated under W,
VA, VAR resolution and
accuracy

± 5 % ± n x 2 % of meas
± 10 counts

Phase between In - Vn

-360º ... +360º

± n × 1.5º

Frequency

0 ... 3500 Hz

1 Hz

± 1Hz

THD

0.0 ... 100.0 %

0.1%

± 5 %

DC relative

absolute

0.0 ... 100.0 %

0.0 ... 100.0 V

0.1 %

0.1 V

± 2 %

± 5 % of meas ± 10
counts