Fluke 5725A User Manual

Page 38

Advertising
background image

5725A

Instruction Manual

1-22

Table 1-15. DC Current Secondary Performance Specifications and Operating Characteristics

Temperature

Coefficient [Note 2]

Noise

Range

Stability

±

1 °C [Note 1]

10 °-40 °C

0 °-10 °C

and

Compliance

Limits

Burden

Voltage

Adder

[Note 3]

Maximum

Load for

Full

Accuracy

Bandwidth

0.1-10 Hz

Bandwidth

10 Hz-10 kHz

24 Hours

40 °-50 °C

[Note 4]

pk-pk

RMS

±

(ppm output

+ nA)

±

(ppm output + nA)/ °C

±

nA/V

ppm output

+ nA

nA

220

µ

A

2.2 mA
22 mA
220 mA
2.2 A

0

5 + 1

0

5 + 5

0

5 + 50

0

8 + 300

0

9 + 7

µ

A

0

1 + 0.40

0

1 + 2

0

1 + 20

0

1 + 200

0

1 + 2.5

µΑ

0

3 + 1

0

3 + 10

0

3 + 100

0

3 + 1

µ

A

0

3 + 10

µ

A

10
10
10
10

3 [Note 5]

0.2
0.2
10
100
2

µ

A

20k
2k
200
20
2

0

6 + .9

0

6 + 5

0

6 + 50

0

9 + 300

12 + 1.5

µ

A

10
10
50
500
20

µ

A

5725A

±

(ppm output

+

µ

A)

±

(ppm output +

µ

A)/

°C

ppm output

+

µ

A

µ

A

11 A

25 + 100

20 + 75

30 + 120

4

0

4

15 + 70

175

Notes:

Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA
ranges are increased by a factor of 1.3 when supplied through 5725A terminals.

1. Stability specifications are included in the Absolute Uncertainty values for the primary specifications.

2. Temperature coefficient is an adder to uncertainty specifications. It does not apply unless operating more

than

±

5 °C from calibration temperature.

3. Burden voltage adder is an adder to uncertainty specifications that does not apply unless burden voltage is

greater than 0.5 V.

4. For higher loads, multiply uncertainty specification by:

1

0.1 x actual load

maximum load for full accuracy

+

5. The calibrator’s compliance limit is 2 V for outputs from 1 A to 2.2 A. 5725A Amplifier may be used in range-

lock mode down to 0 A.

Minimum output: 0 for all ranges, including 5725A.

Settling time to full accuracy: 1 second for mA and mA ranges; 3 seconds for 2.2 A range; 6 seconds for 11 A
range; + 1 second for range or polarity change

Overshoot: <5 %