Operating Notes

5725A Operating Functions and Modes

3

3-11

3-9. Current

Standby

CURRENT

OUTPUT

HI

LO

VOLTAGE MODE

CURRENT MODE

5725A STANDBY

FAULT

OVERLOAD

CABLE OFF

11A

MAX

2OV

PK

MAX

NO OUTPUT

HI

HI

LO

LO

AUX

CURRENT

GUARD

GROUND

WIDEBAND

SENSE

V A

OUTPUT

V A

µ

NO

OUTPUT

HI

BOOST

LIT

aq10i.eps

Figure 3-5. Current Standby Mode

Current standby mode is when both CURRENT MODE and 5725A STANDBY
indicators are lit. In current standby mode, the 5725A is not supplying an output. Current
standby occurs under any of the following conditions:

When the 5700A current output location is set to "5725A", and any ac or dc current
level is showing on the 5700A Output Display with the STANDBY annunciator lit.

If "5725A" is selected in a 5700A setup menu as the current amplifier, and any of the
following conditions are true:

1.

An ac or dc current 2.2A or above is showing on the 5700A Output Display with
the STANDBY annunciator lit.

2.

Any level of dc current is showing on the 5700A Output Display with the
STANDBY annunciator lit, and the BOOST key on the 5700A is pressed so that
the BOOST indicator is lit.

3.

An ac current 1A or above is showing on the 5700A Output Display with the
STANDBY annunciator lit, and the BOOST key on the 5700A is pressed so that
the BOOST indicator is lit.

4.

The 5700A is ranged locked on the 11A range, and any dc current or an ac
current 1A or above is showing on the Output Display with the STANDBY
annunciator lit.